Carbon Compound Instruments
​​​​​​​​​​​​​​​​​​​​​ Conta​​​ct us
Traini​ng
Sales Contacts
Request a Quote
Terms & Conditions
​​ ​
​​

​Model N360M
Mid-Range Gas Filter Correlation CO2 Analyzer
with NumaView™ Software​
N360M_Straight_tdylogo_WEB.png

 Downloads

  
  
  
092910330B - MANUAL, USER, N360, N360M.pdfN360/N360M with NumaView™ Software Manual13697 KB
NVS Remote Manual - 08492.pdfNumaView™ Software Remote User Guide867 KB
SAL000076K - Environmental Instruments Overview.pdfEnvironmental Instruments Overview610 KB
SAL000123A - N360M.pdfN360M Specifications972 KB
Teledyne-N-Series-CEMS-Brochure-Final-01-14-21.pdfN Series Instruments for Continuous Emissions Monitoring (CEM) Brochure11049 KB

 Applications

 Specifications


Ranges
Min: 0 - 4 ppm full scale
Max: 0 - 4,000 ppm full scale (selectable, dual-range supported)
Measurement Units ppb, ppm, μg/m3, mg/m3 (selectable)
​Response Time
​< 70 seconds to 95%
​Sample Flow Rate
​​800 cc/min ±10%
Zero Noise​
< 0.2 ppm (RMS)
Span Noise < 1% of reading (RMS)
Lower Detectable Limit < 0.4 ppm
​Precision
0.5% of reading
​Linearity
1% of full scale
Zero Drift
< 0.​5 ppm/24 hours
Span Drift < 0.5% of reading/24 hours​​​
Included I/O 1 x Ethernet (TCP/IP)
1 x RS232
2 x front panel USB device ports​
Optional I/O Universal Analog Output Board includes (all user-definable):
4 x Isolated Voltage Outputs (5V, 10V; user-selectable)
3 x Individually Isolated Current Outputs (4-20mA)
​Digital I/O Expansion Board includes:
3 x Isolated Digital Input Controls
5 x Isolated Digital Output Controls (user-definable)
3 x Form C Relay Alarm Outputs (user-definable)​
​​Weight
​40 lbs (18.1 kg)
Dimensions (HxWxD)
7” x 17” x 24.3” (178 x 432 x 617 mm)
​​Operating Temperature
​0 - 40ºC
Power

Specifications are subject to change without notice.
All sp​ecifications are based on constant conditions.​
​100V-240V, 50/60 Hz, Typical consumption <150W




 Features

  • Single or dual range capability
  • Customizable alerts and continuous self-checking
  • Wide operating temperature range
  • Internal DC-powered va​cuum pump
  • 5-year warranty on Gas Filter Correlation (GFC) wheel
  • Integrated Oxygen (O2) sensor (optional)​

 Description

The Model N360M measures carbon dioxide (CO2) by comparing infrared energy absorbed by a sample to that absorbed by a reference according to the Beer-Lambert law.

The N360M uses Gas Filter Correlation (GFC) to overcome the interfering effects of various other gases (such as water vapor) that also absorb IR. The analyzer passes the IR beam through a spinning wheel made up of two separate chambers: one containing a high concentration of CO2 known as the reference, and the other containing a neutral gas known as the measure. The concentration of CO2 in the sample chamber is computed by taking the ratio of the instantaneous measure and reference values and then compensating the ratio for sample temperature and pressure. A nitrogen purge system is provided for the GFC wheel assembly to eliminate the eff​ects of ambient CO2, if necessary.

Instrument functions and controls are managed through a series of integrated microprocessor-controlled modules utilizing a simple and reliable CAN Bus communications architecture. Each module is independently assembled and calibrated allowing easy and fast field replacement to maximize instrument uptime.

Intuitive operation and calibration of all N Series products is achieved through the NumaView™ Software interface. The graphical user interface (GUI) is customizable, giving the user fast and efficient access to instrument status, as well as measurement data and diagnostic parameters in either numeric or graphical form. NumaView™ Remote Software (included at no charge) provides the same virtual interface and complete instrument control, as well as access to the instrument’s large internal data storage buffer from a remote PC or tablet.​